产品名称: 校准套件
产品型号: OEM LV 124 AND OEM LV 148
产品展商: EM TEST
产品文档: 无相关文档

简单介绍

The CA LV 124 set includes four load resistors for the verification of the pulses E10 and E13 of the LV 124 (2013). The loadresistors CA LV124-P1R and CA LV124-P100R are used for power line verification. For verification the 16 signal- and datalines the CA LV124-D1R and CA LV124-D1000R are used.


校准套件  的详细介绍

The CA LV 124 set includes four load resistors for the verification of the pulses E10 and E13 of the LV 124 (2013). The loadresistors CA LV124-P1R and CA LV124-P100R are used for power line verification. For verification the 16 signal- and datalines the CA LV124-D1R and CA LV124-D1000R are used.

CALIBRATION SET FOR OEM LV 124 AND OEM LV 148
  • 应用领域
  • 符合下列标准
  • 特点
  • 优点
  • 软件
Data Sheet CA LV 124 (20140704)_201807121317101.jpg
  • 应用领域

    The CA LV 124 set includes four load resistors for the verification of the pulses E10 and E13 of the LV 124 (2013). The load

    resistors CA LV124-P1R and CA LV124-P100R are used for power line verification. 

    For verification the 16 signal- and datalines the CA LV124-D1R and CA LV124-D1000R are used.

  • 符合下列标准

    TEST PROCEDURE

    One reference measurement each with 100 ohm (±5%)

    and 1 ohm (±5%) as a DUT substitute must be performed

    and documented. Verification of the edge steepness must

    be provided with this test setup. Low-inductance parts

    must be used as resistors.

    The load resistor must have the following characteristics

    to achieve the required pulse parameters:

    - Low Inductive Component

    - Shortest possible cables (a few cm) to the PFM output

    - Probe must be as near as possible to the load resistor or,

    as with the CA LV 124-P, using a BNC connection directly

    on the load.

  • 特点

    > Low inductive load resistor for direct connect into

    the PFM 200N100 output power plugs

    > Load resistor 1.0 ohm and 100 ohm for power

    lines verification

    > Load resistor 1.0 ohm and 1000 ohm for datalines

    verification

    > High precision load resistors 1.0, 100 and 1000

    ohms

    > Software procedure for verification with CA LV124

    loads

  • 优点

    > Low inductive load resistor for direct connect into

    the PFM 200N100 output power plugs

    > Load resistor 1.0 ohm and 100 ohm for power

    lines verification

    > Load resistor 1.0 ohm and 1000 ohm for datalines

    verification

    > High precision load resistors 1.0, 100 and 1000

    ohms

    > Software procedure for verification with CA LV124

    loads

  • 软件

    POWER LINE SETTINGS

    The reference measurement can be selected by selecting

    Edge Verification in the software. There is a test window

    for call up the reference measurement settings.

    1 Edge Verification: 

    Choose this option to perform the reference

    measurement.

    2 Test Window for the reference measurement